Common Features
- Operating Modes: Pneumatically driven & electronically controlled, closed, semi-closed and semi-open
- Pneumatically driven electrically controlled ventilator, can be used for emergency clinical resuscitation and respiratory therapy
- Non hazardous environmentally friendly frame, precise construction, elegant design, equipped with automated self-checking function during startup and an automatic self-calibrating sensor
- Display: Built-in 10.4 inch high visibility color TFT display, able-simultaneously display Pressure-Time and Frequency-Time waveforms in real time
- Sigh: Ability to insert 1-8 sigh breaths in every 100 breaths, ventilation should be no less than 1.5 the tidal volume
- Equipped with infrared turbine flow sensor; precisely measures gas flows, stable and reliable
- Power Failure Support Functions: Automatically converts-standby mode during AC power outages
- 9 Level adjustable humidifier
- Monitor Parameters: tidal volume, respiration rate, airway pressure, I:E ratio, inspiratory & expiratory status, inspiratory trigger and others
- Alarms: Audible and/or visual alarms for ventilation volume, tidal volume, oxygen concentration, asphyxia, airway pressure, intubation disconnection, low oxygen, power failure and other alarms, also features a alarm history query function
Advanced Model Exclusive Features
- Equipped with professional medical grade air compressor tailored for long clinical times and low noise requirements
Air Compressor
- Air Compressor: 220V AC±10%, 50±1Hz, 700VA, also features good protective grounding against electrical activity
- Output Gas Pressure: 0.4MPa±20%
- Continuous output gas flow greater than or equal to 20L/min
- Instantaneous output gas flow great than equal to 60L/min
Model |
MKR.VM700B (Standai'd Model) |
MKR.VM700B (Advanced Model) |
Display |
10.4 inch high visibility color TFT display |
Ventilation Mode |
A C, IPPV, SIPPV, IMV SIMV, Standby, SPONT MANUAL |
Minute Vblume |
Greater than or equal to 18L |
Back-up Power Supply |
At least 4 hours |
Tidal blume |
Adjustable range: 50-1500mL Display range: 0-2000ml |
Respiratoiy Rate |
l-99bpm |
Inspiratoiy expiratoiy (I:E) Ratio |
4:1-1:4 |
PEEP |
O-lOcmH^O £ |
SIMV Rate |
l-20bpm |
Inspiratory Pressure Trigger |
-10-10cmH^O £ |
Inspiratory Plateau |
Adjustable fi'om 0-50% of inspiratory time |
Adjustable Oxygen Concentration |
1. 45-100% 2. Without air compressor |
1. 21-100% 2. With air compressor |
Gas Supply Requirement |
280 ~ 600 kPa medical grade oxygen and compressed air supply |
Pressure Limit |
l.OkPa 〜6.0kPa |
Maximum Safe Operating Pressure |
<6.0kPa |
Tidal Volume Alarm |
Upper Limit: 10-2000mL Lower Limit: 0-1800ml |
Ainvay Pressure Alarm |
Upper Limit: 0.1 〜6kPa, Lower Limit: 0〜5kPa |
Oxygen Concentration Alarm |
Upper Limit: 22-100%. Lower Limit: 21-80%
|